In situ calibration of the foil detector for an infrared imaging video bolometer using a carbon evaporation technique
この論文をさがす
抄録
0000-0003-1586-1084
The InfraRed imaging Video Bolometer (IRVB) is a useful diagnostic for the multi-dimensional measurement of plasma radiation profiles. For the application of IRVB measurement to the neutron environment in fusion plasma devices such as the Large Helical Device (LHD), in situ calibration of the thermal characteristics of the foil detector is required. Laser irradiation tests of sample foils show that the reproducibility and uniformity of the carbon coating for the foil were improved using a vacuum evaporation method. Also, the principle of the in situ calibration system was justified.
収録刊行物
-
- Review of Scientific Instruments
-
Review of Scientific Instruments 87 11E124-, 2016-11-11
AIP Publishing