心磁図のICAを用いた環境磁気ノイズ除去におけるパラメータ決定の検討  [in Japanese] Study of Parameter Decision for a Noise Rejection Method Using ICA for a Magnetocardiogram  [in Japanese]

Access this Article

Search this Article

Author(s)

Abstract

  We studied an environmental magnetic noise rejection method using independent component analysis (ICA) for a magnetocardiogram (MCG). In the ICA, when the SN ratio of the data is reduced, the accuracy of the noise rejection becomes low and it becomes difficult to reach a decision regarding the contraction dimension. In order to solve these problems, we examined the initial value of the transformation matrix W of the ICA. We proposed an initial value decision procedure for the transformation matrix W, and as a result the accuracy of the noise rejection was improved. Verification was carried out for MCGs with various SN ratios. It was shown that this method was effective for improving noise rejection.

Journal

  • Journal of the Magnetics Society of Japan

    Journal of the Magnetics Society of Japan 34(2), 146-150, 2010

    The Magnetics Society of Japan

Codes

  • NII Article ID (NAID)
    130000168431
  • NII NACSIS-CAT ID (NCID)
    AA12297999
  • Text Lang
    JPN
  • ISSN
    1882-2924
  • NDL Article ID
    10610120
  • NDL Source Classification
    ZM35(科学技術--物理学)
  • NDL Call No.
    Z15-398
  • Data Source
    NDL  J-STAGE 
Page Top