Irradiation Experiments on a Mouse Using a Mild-Plasma Generator for Medical Applications

  • SAKAKITA Hajime
    Energy Technology Research Institute, National Institute of Advanced Industrial Science and Technology (AIST), Tsukuba Central 2, 1-1-1 Umezono, Tsukuba 305-8568, Japan
  • IKEHARA Yuzuru
    Research Center for Medical Glycoscience, National Institute of Advanced Industrial Science and Technology (AIST), Tsukuba Central 2, 1-1-1 Umezono, Tsukuba 305-8568, Japan

抄録

Plasma technologies using an argon plasma coagulator have been used in endoscopic therapy to induce blood coagulation and ablate residual tumors. However, present devices have a risk of perforating the stomach wall during endoscopic submucosal dissection. Therefore, to reduce this risk, irradiation is performed for a limited time, which leads to incomplete cessation of bleeding and recurrence of residual tumors. Therefore, a device with greater controllability and safety is strongly desired for clinical applications. In this study, we have evaluated the irradiation efficiency of an atmospheric-pressure plasma jet based on a dielectric barrier discharge to control bleeding. Bleeding from a mouse femoral artery was induced, and then plasma was irradiated onto the bleeding area. Prompt coagulation in the disrupted blood vessel was observed, and there was no histological evidence of either burns or tissue necrosis caused by the plasma jet. These results suggest that postoperative scarring and adhesion may be prevented using the proposed plasma generator because of the reduced tissue damage.

収録刊行物

被引用文献 (18)*注記

もっと見る

参考文献 (5)*注記

もっと見る

詳細情報 詳細情報について

問題の指摘

ページトップへ