グルタルアルデヒド処理ウマ心膜を用いた自作valved conduitの弁機能に関する実験的検討
書誌事項
- タイトル別名
-
- Experimental study of valvular function of valved conduit using glutaraldehyde-preserved equine pericardium(Xenomedica).
抄録
Hydrodynamics of valved conduit made of equine pericardium (Xenomedica®) was studied. Monocusp, bicusp and tricusp valved conduits were evaluated for valvular function by original circulation system. The study was also performed with the conduit being compressed. Flow rate curve and flow-pressure gradient Lissajous were useful for evaluation. Conduit with monocusp valve had diastolic regurgitation. Tricusp-valved conduit had increased valvular registance. On the other hand, bicusp-valved conduit had stable performance under any condition.
収録刊行物
-
- 人工臓器
-
人工臓器 17 (3), 1121-1124, 1988