グルタルアルデヒド処理ウマ心膜を用いた自作valved conduitの弁機能に関する実験的検討

DOI

書誌事項

タイトル別名
  • Experimental study of valvular function of valved conduit using glutaraldehyde-preserved equine pericardium(Xenomedica).

抄録

Hydrodynamics of valved conduit made of equine pericardium (Xenomedica®) was studied. Monocusp, bicusp and tricusp valved conduits were evaluated for valvular function by original circulation system. The study was also performed with the conduit being compressed. Flow rate curve and flow-pressure gradient Lissajous were useful for evaluation. Conduit with monocusp valve had diastolic regurgitation. Tricusp-valved conduit had increased valvular registance. On the other hand, bicusp-valved conduit had stable performance under any condition.

収録刊行物

  • 人工臓器

    人工臓器 17 (3), 1121-1124, 1988

    一般社団法人 日本人工臓器学会

詳細情報 詳細情報について

  • CRID
    1390282679937330560
  • NII論文ID
    130001734348
  • DOI
    10.11392/jsao1972.17.1121
  • ISSN
    18836097
    03000818
  • 本文言語コード
    ja
  • データソース種別
    • JaLC
    • CiNii Articles
  • 抄録ライセンスフラグ
    使用不可

問題の指摘

ページトップへ