Hollow Fiber型人工肺の臨床経験  coil型人工肺との比較

DOI

書誌事項

タイトル別名
  • Clinical experience of Hollow Fiber type Oxygenator (BOS CM40).
  • In Comparison with Coil Type Membrane Oxygena t or (Koloboes 3500-IIA)
  • ―Coll型人工肺との比較―

抄録

The Hollow Fiber Oxygena t or (BOS CM40) was compared with coil type membrane oxygenator (Kolobow-3500IIA) in 42 patients who underwent valve replacement. There was no statistically significant difference in platelet count plasma free hemoglobin levels and inmunoglobulin levels between the both oxygenators. No marked increment of blood fibrinolysis activity was found throughout cardiopulmonary (CP) bypass in the both. Rapid reduction in the levels of CH50, C3 and C4 within five minutes after the initiation of CP bypass indicated rapid complement activation via the classical pathway. The level of C3a showed a progressive increase during CP bypass in the both, especially in Kolobow 3500-IIA demonstrating an early significant (p<0.05) elevation of C3a activity. As far as complement activation is concerned, BOS CM40 seems to be superior to Kolobow 3500-IIA.

収録刊行物

  • 人工臓器

    人工臓器 16 (1), 625-628, 1987

    一般社団法人 日本人工臓器学会

詳細情報 詳細情報について

  • CRID
    1390001204959247232
  • NII論文ID
    130001735187
  • DOI
    10.11392/jsao1972.16.625
  • ISSN
    18836097
    03000818
  • 本文言語コード
    ja
  • データソース種別
    • JaLC
    • CiNii Articles
  • 抄録ライセンスフラグ
    使用不可

問題の指摘

ページトップへ