Mitroflow人工弁の臨床的機能評価

DOI

書誌事項

タイトル別名
  • Clinical evaluations of Mitroflow bovine pericardial valve.

この論文をさがす

抄録

For the purpose of clinical evaluations of the Mitroflowvalve, we have measured effective valvuler orifice area (EOA), cardiac output at the heart rate of 90-180bpm made by the atrial pacing, and cineangiographic studies of cusp motion. The results were as follows: EOA were 2.28 (21A), 2.77 (23A), 3.12 (27M), 3.24, 3.52, 3.60cm2 (29M). The cardiac output increased up to 170bpm in 2 cases studied. Cineangiographic studies of the cusp movement showed synergic opening and closing of the 3 cusps and maximum opening of each cusp at mid diastole or systole.

収録刊行物

  • 人工臓器

    人工臓器 14 (3), 1298-1301, 1985

    一般社団法人 日本人工臓器学会