血液ろ過における循環血液量の自動制御:循環血液量変化の測定方法と制御方法
書誌事項
- タイトル別名
-
- Automatic control of circulating blood volume during hemofiltration
抄録
A program control system of circulating blood volume during hemofiltration has been developed, which is constructed from a continuous circulating blood volume measuring system, a micro-computer, and a substitution fluid infuser. In this system, hematocrit is calculated from the formulae as a function of electrical resistivities of blood and filtrate, which was given by Geddes and Gram. Then the circulating blood volume change, ΔBV, is calculated and the flow rate of the infuser is adjusted by the micro-computer. This system was applied to 8 patient in a total of 30 times and was found to keep the difference between actual and desired blood volume values within ±3%.
収録刊行物
-
- 人工臓器
-
人工臓器 13 (2), 724-727, 1984