筋機能MR画像の収集時間短縮化の試み
書誌事項
- タイトル別名
-
- Reduction of acquisition time for muscle functional magnetic resonance images
この論文をさがす
抄録
Exercise induced muscle activity is essential in sports medicine. The previous studies proposed muscle functional magnetic resonance imaging (mfMRI) which visualized muscle activity. However, the mfMRI used the spin echo (SE) sequence, and SE requires several minutes for acquisition time. The purpose of this study was to judge the feasibility of the mfMRI using ultrafast imaging (fast-acquired mfMRI: fast-mfMRI) . The 10 subjects performed 200 repetitions of ankle plantar flexion. T2 relaxation curves of muscle were compared with multiple spin echo (MSE) and spin-echo echo-planar-imaging (SE-EPI) . The fast-mfMRI was the images which the processed images fused into the morphologic images. In the result, there was no problem using the T2 relaxations MR signals from SE-EPI in the range of the short TE. In the fused images, the areas of activated muscle in the fast-mfMRI from the SE-EPI images were in good agreement with that of the fast-mfMRI from the MSE images. In the geometric location of the fast-mfMRI, there was no major difference between the morphologic image and the functional image. Morphological and functional information from the fast-mfMRI can be applied to the human trunk which has a limited scan time.
収録刊行物
-
- 医用画像情報学会雑誌
-
医用画像情報学会雑誌 26 (1), 1-7, 2009
医用画像情報学会
- Tweet
キーワード
詳細情報 詳細情報について
-
- CRID
- 1390282679628700032
-
- NII論文ID
- 130002069144
-
- ISSN
- 18804977
- 09101543
-
- 本文言語コード
- ja
-
- データソース種別
-
- JaLC
- CiNii Articles
- KAKEN
-
- 抄録ライセンスフラグ
- 使用不可