高速液体クロマトグラフィーによる血清中チオペンタール濃度測定法の臨床応用

書誌事項

タイトル別名
  • The Clinical Application of High-Performance Liquid Chromatographic Method for Thiopental Determination in Human Serum.

この論文をさがす

抄録

It can be presumed that liver function will be influenced by the administration of thiopental in the course of barbiturate therapy.Fof this reason, we established and tested a new method for the therapeutic monitoring of the concentration of thiopental in human serum.<BR>In this method, we employed high-performance liquid chromatography (HPLC) to simplify the pretreatment technique, thereby to shorten the amount of time needed for measurement. As a result, chromatographic data, and the calibration curve (Y=0.249X+0.118, r=0.999) obtained indicated that within-, between-run precision, and recovery rates from the serum are all extremely good. The lowest measurable concentration was found to be 0.1μg/ml (Aufs 0.16). Also, in the cross-reaction tests among drugs mainly used in combination, we did not find any influence on the quantitative analysis of thiopental. We, then measured the serum levels in patients to whom thiopental was administered intravenously. By submitting the results of the t-test, we confirmed a significant rise in the concentration of thiopental in patients with the liver failure, as compared with normal patients (p<0.01). Therefore, we conclude that this assay system can provide measurement as fast as possible, with the added benefit of simplified clinical procedures.

収録刊行物

  • 病院薬学

    病院薬学 18 (1), 72-78, 1992

    一般社団法人 日本医療薬学会

詳細情報 詳細情報について

問題の指摘