Signal Enhancement by Crossing the Sample Flow at a Small Angle against the Laser Beam in Multiphoton Ionization Mass Spectrometry

DOI PDF Web Site Web Site Web Site View 1 Remaining Hide 5 Citations 21 References
  • MATSUI Taiki
    Department of Applied Chemistry, Graduate School of Engineering, Kyushu University
  • IMASAKA Totaro
    Department of Applied Chemistry, Graduate School of Engineering, Kyushu University Division of Optoelectronics and Photonics, Center for Future Chemistry, Kyushu University

Search this article

Abstract

To enhance sensitivity in multiphoton ionization/time-of-flight mass spectrometry, the sample, in the form of an effusive molecular flow, was introduced at a small angle relative to the laser beam. This approach enlarged the volume of interaction between the sample and the laser beam, thus increasing the number of ions formed by multiphoton ionization. The sensitivity obtained using this technique was compared with that obtained using a previous method based on the off-axis sample introduction technique with detection limits for organic hydrocarbons at the subfemtogram level. The detection limit was examined for dioxins and an improvement of 2.5 ± 0.4 fold was found. The technique developed herein is applicable for practical ultratrace analysis of persistent organic pollutants.

Journal

  • Analytical Sciences

    Analytical Sciences 30 (4), 445-449, 2014

    The Japan Society for Analytical Chemistry

Citations (5)*help

See more

References(21)*help

See more

Related Projects

See more

Details 詳細情報について

Report a problem

Back to top