Study on the Development of a Patient Dosimetry Gown for Interventional Cardiology Procedures
-
- Kato Mamoru
- Department of Medical Technology, Akita Medical Center Tohoku University Graduate School of Medicine
-
- Chida Koichi
- Tohoku University Graduate School of Medicine
-
- Moritake Takashi
- Institute of Industrial Ecological Sciences, University of Occupational and Environmental Health
-
- Koguchi Yasuhiro
- Oarai Research Center, Chiyoda Technol Corporation
-
- Kaga Yuji
- Department of Radiology, Sendai Kosei Hospital
-
- Sakamoto Hajime
- Department of Radiology, University of Yamanashi Hospital
-
- Tsukamoto Atsuko
- Department of Radiology, Kanto Medical Center NTT EC
-
- Kawauchi Satoru
- Department of Radiological Technology, Toranomon Hospital
-
- Matsumoto Kazuma
- Department of Radiology, Hyogo College of Medicine Hospital
-
- Matsumura Mitsuaki
- Department of Radiological Technology, Japanese Red Cross Kobe Hospital
-
- Oosaka Hajime
- Department of Medical Technology, Akita Medical Center
-
- Tosa Tetsuo
- Department of Medical Technology, Akita Medical Center
Bibliographic Information
- Other Title
-
- 心血管撮影領域における患者被ばく線量測定用装具の開発
- シンケッカン サツエイ リョウイキ ニ オケル カンジャ ヒバクセンリョウ ソクテイヨウ ソウグ ノ カイハツ
Search this article
Abstract
In recent years, dose justification and optimization have been attempted in percutaneous coronary intervention (PCI); however, deterministic effects have been reported. To prevent radiation skin injuries in PCI, it is necessary to measure the patient entrance skin dose (ESD), but an accurate dose measurement method has not yet been established. In this study, we developed a dosimetry gown that can measure the ESD during PCI using multiple radiophotoluminescence dosimeters (RPLDs). The RPLDs were placed into 84 pockets that were sewn into a dosimetry gown. Patients wear the original dosimetry gown during the procedures, after which we obtain accurate ESD measurements. We believe that this method using RPLDs and a newly-designed dosimetry gown provides accurate ESD measurements during PCI. We expect this system to become a standard method for measuring ESD during PCI.
Journal
-
- Japanese Journal of Radiological Technology
-
Japanese Journal of Radiological Technology 70 (8), 814-820, 2014
Japanese Society of Radiological Technology
- Tweet
Keywords
Details 詳細情報について
-
- CRID
- 1390001206363597056
-
- NII Article ID
- 130004678739
-
- NII Book ID
- AN00197784
-
- COI
- 1:STN:280:DC%2BC2M%2FjvVKgtQ%3D%3D
-
- ISSN
- 18814883
- 03694305
-
- NDL BIB ID
- 025779172
-
- PubMed
- 25142393
-
- Text Lang
- ja
-
- Data Source
-
- JaLC
- NDL
- Crossref
- PubMed
- CiNii Articles
- KAKEN
-
- Abstract License Flag
- Disallowed