Temporal Changes of Fractionation Index Caused by Changes in the Large Size of Ablated Particles in Laser Ablation–Inductively Coupled Plasma Mass Spectrometry

DOI PDF Web Site Web Site Web Site View 1 Remaining Hide 6 Citations 23 References Open Access
  • MACHIDA Ryo
    Faculty of Science and Engineering, Department of Applied Chemistry, Chuo University
  • NAKAZAWA Takashi
    Faculty of Science and Engineering, Department of Applied Chemistry, Chuo University
  • FURUTA Naoki
    Faculty of Science and Engineering, Department of Applied Chemistry, Chuo University

Search this article

Abstract

To elucidate mechanisms of elemental fractionation that are observed in laser ablation–inductively coupled plasma mass spectrometry, the relative intensities of 34 elements, each normalized by a Ca internal standard, were measured every minute during a 10-min laser ablation of an NIST 610 glass standard. Temporal changes in the fractionation index (FI) were obtained by dividing the relative intensity of every minute by that of the first minute. The particles generated by laser ablation were collected on a filter every minute, and they were observed using scanning electron microscopy to investigate changes in the large size of particles. Large variations among the large size of particles were observed using single-site mode and under 1.0 mm defocus conditions. The 34 measured elements were classified into two groups, depending on their observed FI variation. The FI variation was rationalized by elemental behavior due to changes in the large size of ablated particles introduced into the ICP.

Journal

  • Analytical Sciences

    Analytical Sciences 31 (5), 345-355, 2015

    The Japan Society for Analytical Chemistry

Citations (6)*help

See more

References(23)*help

See more

Related Projects

See more

Details 詳細情報について

Report a problem

Back to top