Complementary Use of LC-ICP-MS and LC-ESI-Q-TOF-MS for Selenium Speciation

DOI PDF Web Site Web Site Web Site ほか1件をすべて表示 一部だけ表示 被引用文献6件 参考文献17件
  • ANAN Yasumi
    Laboratory of Chemical Toxicology and Environmental Health, Showa Pharmaceutical University
  • NAKAJIMA Genki
    Laboratory of Chemical Toxicology and Environmental Health, Showa Pharmaceutical University
  • OGRA Yasumitsu
    Laboratory of Chemical Toxicology and Environmental Health, Showa Pharmaceutical University Department of Toxicology and Environmental Health, Graduate School of Pharmaceutical Sciences, Chiba University

この論文をさがす

抄録

We demonstrated the complementary use of inductively coupled plasma–mass spectrometry (ICP-MS) and electrospray ionization quadrupole time-of-flight mass spectrometry (ESI-Q-TOF-MS) for the analysis of Se-containing compounds, such as selenate, selenomethionine (SeMet), and trimethylselenonium ion (TMSe), found in biological samples. The sensitivity of ESI-Q-TOF-MS for Se-containing compounds was strongly dependent on the chemical species. ICP-MS exhibited higher sensitivity than ESI-Q-TOF-MS, and had no species dependency. On the other hand, ESI-Q-TOF-MS enabled easy and robust identification of Se-containing compounds.

収録刊行物

  • Analytical Sciences

    Analytical Sciences 31 (6), 561-564, 2015

    社団法人 日本分析化学会

被引用文献 (6)*注記

もっと見る

参考文献 (17)*注記

もっと見る

詳細情報 詳細情報について

問題の指摘