Sequential Analysis of β-Lactoglobulin for Allergen Check Using QCM with a Passive Flow System
-
- Takeshi Ito
- Kanagawa Industrial Technology Center
-
- Nobuyoshi Aoki
- Kanagawa Industrial Technology Center
-
- Akihisa Tsuchiya
- Kanagawa Industrial Technology Center
-
- Satoru Kaneko
- Kanagawa Industrial Technology Center
-
- Koji Suzuki
- Keio University
抄録
<jats:title>Abstract</jats:title> <jats:p>Quantitative detection of food allergen is placed considerable emphasis in food factories all over the world. We proposed a quartz crystal microbalance (QCM) method coupled with a passive flow system for the analysis of β-lactoglobulin, as an example. Analytical curves for a passive were comparable on an active flow system, and 1 ppm of β-lactoglobulin was detected. Repeatable detection was achieved by the regeneration of the sensor surface. One detection cycle was completed within 13 min. This proposed system can be used for quantitative analysis without professional skills, only by dropping a liquid sample.</jats:p>
収録刊行物
-
- Chemistry Letters
-
Chemistry Letters 44 (7), 981-983, 2015-05-12
Oxford University Press (OUP)