Non-restriction Monitoring of Apnea-hypopnea by Oral Throat Sound Measurement System
-
- Uraminami Shoichi
- Osaka Electro-Communication University Graduate School
-
- Sakai Noriaki
- Department of Clinical Engineering, Daini Okamoto General Hospital
-
- Oishi Toshimitsu
- Osaka Electro-Communication University Graduate School
-
- Matsumura Masafumi
- Osaka Electro-Communication University Graduate School
Bibliographic Information
- Other Title
-
- 口腔咽喉音分析に基づく無呼吸・低呼吸の無拘束モニタリング
Abstract
In this study, a breath monitoring system is being developed using a contact type microphone attached the neck and measuring the sound at the time of respiration (breath sound). In this paper, I made s sound measuring amplifier with adjustable frequency bandwidth amplification degree, so it fits each subject properly as well as improving the sound measuring circuit. Then I demonstrated that normal breathing, apnea and hypopnea could be determined analysing the relation between nasal airflow and the amplitude of the sound produced at neck level doing respiration. In addition, I verified the efficacy of the breath monitoring system by comparing it to an equipment that can, in a simple manner, examine the sleep apnea.
Journal
-
- IEEJ Transactions on Electronics, Information and Systems
-
IEEJ Transactions on Electronics, Information and Systems 136 (7), 1011-1012, 2016
The Institute of Electrical Engineers of Japan