Quantitative Analysis of Trace Elements in Silicate Glass Sample by LA-ICP-QMS/QMS with an ORC: Silicon as the Matrix of Calibrating Solutions and the Internal Standard for Measurement

DOI PDF Web Site Web Site Web Site ほか1件をすべて表示 一部だけ表示 被引用文献3件 参考文献27件
  • ZHU Yanbei
    National Metrology Institute of Japan, National Institute of Advanced Industrial Science and Technology

この論文をさがす

抄録

A strategy was presented for quantitative analysis of trace elements in a glass sample by laser ablation inductively coupled plasma tandem quadrupole mass spectrometry (LA-ICP-QMS/QMS) with an octapole reaction cell (ORC). Silicon in the glass was used as the internal standard as well as the matrix for making the calibrating solutions. Sample aerosols generated by the laser ablation system were introduced into the dual pass spray chamber through the make-up gas port. Calibrating solutions were nebulized into the spray chamber using a microflow nebulizer. The silicon matrix-matched calibrating solutions produced a calibration curve capable for the quantitative analysis of trace elements in a silicate glass sample, NIST SRM 612. The analytical results agreed with the certified values, taking into consideration their expanded uncertainties. The detection limits for Cr, Mn, Fe, Ni, Cu, As, Sr, Ag, Cd, Sb, Ba, and Pb, were respectively 0.3, 0.08, 0.5, 0.4, 0.19, 1.1, 0.1, 0.02, 0.03, 0.025, 0.09, and 0.07 μg g−1.

収録刊行物

  • Analytical Sciences

    Analytical Sciences 32 (11), 1237-1243, 2016

    社団法人 日本分析化学会

被引用文献 (3)*注記

もっと見る

参考文献 (27)*注記

もっと見る

詳細情報

問題の指摘