Improvement of the Reynolds Stress Probe for End-Plate Biasing Experiments in a Cylindrical Laboratory Plasma

Abstract

<p>We improved the signal-to-noise ratio of a Reynolds stress (RS) probe in order to measure the Reynolds stress more accurately. By introducing a shield pipe in the probe, the power spectral density of the noise was suppressed to approximately one tenth in an electric biasing experiment. It was also confirmed that the leak current in the RS probe, which had been used in previous experiments, was small enough for the study of drift waves in our linear device.</p>

Journal

  • Plasma and Fusion Research

    Plasma and Fusion Research 11 (0), 1201091-1201091, 2016

    The Japan Society of Plasma Science and Nuclear Fusion Research

References(8)*help

See more

Related Projects

See more

Details 詳細情報について

Report a problem