Trueness and precision of digital impressions obtained using an intraoral scanner with different head size in the partially edentulous mandible

  • Hayama Hironari
    Removable Partial Prosthodontics, Oral Health Sciences, Graduate School of Medical and Dental Sciences, Tokyo Medical and Dental University
  • Fueki Kenji
    Removable Partial Prosthodontics, Oral Health Sciences, Graduate School of Medical and Dental Sciences, Tokyo Medical and Dental University
  • Wadachi Juro
    Removable Partial Prosthodontics, Oral Health Sciences, Graduate School of Medical and Dental Sciences, Tokyo Medical and Dental University
  • Wakabayashi Noriyuki
    Removable Partial Prosthodontics, Oral Health Sciences, Graduate School of Medical and Dental Sciences, Tokyo Medical and Dental University

Search this article

Abstract

<p>Purpose: It remains unclear whether digital impressions obtained using an intraoral scanner are sufficiently accurate for use in fabrication of removable partial dentures. We therefore compared the trueness and precision between conventional and digital impressions in the partially edentulous mandible.</p><p>Methods: Mandibular Kennedy Class I and III models with soft silicone simulated-mucosa placed on the residual edentulous ridge were used. The reference models were converted to standard triangulated language (STL) file format using an extraoral scanner. Digital impressions were obtained using an intraoral scanner with a large or small scanning head, and converted to STL files. For conventional impressions, pressure impressions of the reference models were made and working casts fabricated using modified dental stone; these were converted to STL file format using an extraoral scanner. Conversion to STL file format was performed 5 times for each method. Trueness and precision were evaluated by deviation analysis using three-dimensional image processing software.</p><p>Results: Digital impressions had superior trueness (54–108 μm), but inferior precision (100–121 μm) compared to conventional impressions (trueness 122–157 μm, precision 52–119 μm). The larger intraoral scanning head showed better trueness and precision than the smaller head, and on average required fewer scanned images of digital impressions than the smaller head (p < 0.05). On the color map, the deviation distribution tended to differ between the conventional and digital impressions.</p><p>Conclusions: Digital impressions are partially comparable to conventional impressions in terms of accuracy; the use of a larger scanning head may improve the accuracy for removable partial denture fabrication.</p>

Journal

Citations (9)*help

See more

References(33)*help

See more

Related Projects

See more

Report a problem

Back to top