書誌事項
- タイトル別名
-
- Odor Biosensor System based on Image Lock-in Measurement for Odorant Discrimination
- ニオイ バイオセンサ オウトウ ガゾウ ノ ロックイン ケイソク オ モチイタ ニオイ シキベツ
この論文をさがす
抄録
<p>There has been so far few odor sensors with an array of the biosensor elements. However, an odor sensor constructed in this fashion is expected to have high performance in terms of the ability to discriminate between odorants. In this study, a parallel lock-in measurement circuit was developed to acquire the fluorescent image of the cells even under an ambient light. Since that circuit can extract only the responses of the cells from the fluorescent image, its output represents a specific pattern corresponding to the odorant. Therefore, principal component analysis was applied to that output in order to confirm the availability of the odorant discrimination by using a fluorescent image. It was found that the output pattern from the developed circuit included information enough to discriminate between at least two types of odorants.</p>
収録刊行物
-
- 電気学会論文誌E(センサ・マイクロマシン部門誌)
-
電気学会論文誌E(センサ・マイクロマシン部門誌) 138 (10), 477-484, 2018-10-01
一般社団法人 電気学会