Separation and Preconcentration of Nickel(II) from Drinking, Spring, and Lake Water Samples with Amberlite CG-120 Resin and Determination by Flame Atomic Absorption Spectrometry

DOI PDF Web Site Web Site Web Site ほか1件をすべて表示 一部だけ表示 参考文献32件
  • ARSLAN Yasin
    Nanoscience and Nanotechnology Department, Faculty of Arts & Science, Mehmet Akif Ersoy University
  • KABAK Burcu
    Chemistry Department, Faculty of Arts & Science, Mehmet Akif Ersoy University
  • TRAK Digdem
    Chemistry Department, Faculty of Arts & Science, Mehmet Akif Ersoy University
  • KENDÜZLER Erdal
    Chemistry Department, Faculty of Arts & Science, Mehmet Akif Ersoy University

この論文をさがす

抄録

<p>In this study, Amberlite CG-120 adsorbent was used for the separation/preconcentration of Ni(II) ions in commercial drinking, spring and lake water samples before detection by flame atomic absorption spectrometry. Various optimization parameters for Ni(II) determination, such as pH, eluent type and concentration, sample and eluent flow rates, amount of adsorbent, were investigated to obtain better sensitivity, accuracy, precision and quantitative recovery. Furthermore, the interference effects of some ions on the recovery efficiency of Ni(II) were also investigated. The optimum experimental parameters were obtained in the case of pH 1; 5 mL of 4 mol L−1 HCl for eluent and 0.3 g for the adsorbent amount. The limit of detection was found to be 0.58 μg L−1 and linearity ranged from 5 to 50 μg L−1. The accuracy of the method was tested by the certified reference material of TMDA-70.2 Ontario Lake Water at a 95% confidence level.</p>

収録刊行物

  • Analytical Sciences

    Analytical Sciences 34 (10), 1143-1147, 2018-10-10

    社団法人 日本分析化学会

参考文献 (32)*注記

もっと見る

詳細情報 詳細情報について

問題の指摘

ページトップへ