円錐型ウォームギア機構を用いた腸管自走システムの開発

書誌事項

タイトル別名
  • Development of a Self-Propelled Colonoscope Robot with a Conical Worm Gear Mechanism
  • エンスイガタ ウォームギア キコウ オ モチイタ チョウカンジソウ システム ノ カイハツ

この論文をさがす

抄録

<p>Incidence rate of colon cancer is the highest among all cancers in Japan while the rate is still being on an upward trend. It is generally known that the colon cancer diagnosed in its early stage, is permanently curable by a minimally invasive manner. The first choice of diagnosis for colon cancer is colonoscopy while the inspection requires a great skill, because the manipulation of endoscope surrounded by the layer of soft intestinal tissues is highly cumbersome (e. g. large intestines only stretches and the tip of endoscope does not advance even the proximal end of endoscope is pushed hard by the inspector). In order to solve this problem, we propose a self-propelled colonoscope robot with a conical worm gear mechanism. The drive force is provided from the motor outside the patient body by a torque tube to a conical worm gear to turn four spur gears arranged in a radial fashion at the tip. Thus the robot can move forward and backward in intestine. The notable feature of the mechanism is that the conical worm gear lets the four spur gears moved in a radial fashion, thus the mechanism provides sufficient propelling force even in the intestines with different radii. A prototype mechanism was developed as a proof-of-concept model by using a 3D printer and driven by a motor. The preliminary experiments on a simulated intestine with different radii (ranging from 30 to 50 mm) and in-vitro tissues showed the promising results.</p>

収録刊行物

被引用文献 (1)*注記

もっと見る

参考文献 (15)*注記

もっと見る

関連プロジェクト

もっと見る

詳細情報 詳細情報について

問題の指摘

ページトップへ