A simple determination of whole-rock major- and trace-element composition for peridotite by micro-XRF spectrometer and ICP-MS using fused-glass bead

  • Akizawa Norikatsu
    Atmosphere and Ocean Research Institute, The University of Tokyo
  • Ishikawa Akira
    Department of Earth and Planetary Sciences, Tokyo Institute of Technology
  • Kogiso Tetsu
    Graduate School of Human and Environmental Studies, Kyoto University

Search this article

Abstract

<p>A simple analytical procedure for determination of whole-rock major- and trace-element composition by micro-X-ray fluorescence (μXRF) spectrometer and inductively coupled plasma-mass spectrometer (ICP-MS) using fused-glass bead (sample + lithium tetraborate) is presented. In the case of peridotites, chromian spinel (Cr-spinel) is one of the minerals resistant to acids and interferes with accurate and reproducible determination of whole-rock composition. Such resistant minerals were not observed in the fused-glass beads prepared here, suggesting complete digestion of the samples including Cr-spinels. The μXRF spectrometer was employed to determine SiO2, MgO, and total Fe2O3 contents with the fused-glass beads. They were subsequently dissolved into a nitric acid solution to analyze thirty seven elements, Na, Al, P, K, Ca, Sc, Ti, V, Cr, Mn, Co, Ni, Cu, Rb, Sr, Y, Zr, Nb, La, Ce, Pr, Nd, Sm, Eu, Gd, Tb, Dy, Ho, Er, Tm, Yb, Lu, Hf, Ta, Pb, Th, and U with sector magnetic field ICP-MS (ICP-SFMS) and Q-pole mass filter ICP-MS (ICP-QMS). The analytical procedures were optimized and evaluated with geological reference materials of JP-1 and BIR-1a, and applied to various mantle xenoliths (lherzolite, harzburgite, dunite, wehrlite, olivine clinopyroxenite, and orthopyroxenite) from Tahiti and Moorea islands (Society Archipelago). Since only fused-glass beads are required as an analytical target to determine whole-rock major- and trace-element compositions, the sample amount consumed through the series of analytical procedure can be constrained to a minute amount (e.g., <0.4 g). The analytical procedures presented here are considered appropriate for tiny and/or precious samples, such as mantle xenoliths and rocks collected by dredging, diving, and drilling from the world’s ocean basins.</p>

Journal

Citations (5)*help

See more

References(40)*help

See more

Related Projects

See more

Details 詳細情報について

Report a problem

Back to top