Time-correlated Single Ion Counting Mass Spectrometer with Long and Short Time-of-Flight Tubes and an Evaluation of Its Performance for Use in Trace Analysis of Allergenic Substances

DOI HANDLE PDF Web Site Web Site ほか2件をすべて表示 一部だけ表示 被引用文献3件 参考文献32件 オープンアクセス
  • JU Tiantian
    Division of International Strategy, Center of Future Chemistry, Kyushu University
  • YOSHINAGA Katsunori
    Division of International Strategy, Center of Future Chemistry, Kyushu University Department of Environmental Design, Faculty of Design, Kyushu University
  • IMASAKA Tomoko
    Department of Environmental Design, Faculty of Design, Kyushu University
  • NAKAMURA Hiroshi
    Division of International Strategy, Center of Future Chemistry, Kyushu University
  • IMASAKA Totaro
    Division of International Strategy, Center of Future Chemistry, Kyushu University Hikari Giken, Co.

この論文をさがす

抄録

<p>An analyte molecule was ionized using a femtosecond laser as the ionization source and was measured by a twin-type time-of-flight mass spectrometer with long (42 cm) and short (6.4 cm) flight tubes. The signal was measured using an analog signal digitizer and a time-correlated single ion counting system, and performance was evaluated by comparing data obtained from both instruments. The short mass spectrometer had a mass resolution of 450 and was used in the trace analysis of allergenic substances in a fragrance.</p>