Impact of Valve Plane Alignment on the Repeatability of Left Ventricular Ejection Fraction in ECG-gated Myocardial SPECT Using Corridor 4DM
-
- Rambow Franziska
- Department of Diagnostic and Interventional Radiology and Nuclear Medicine, University Medical Centre Hamburg-Eppendorf
-
- Gersdorf Denis
- Department of Diagnostic and Interventional Radiology and Nuclear Medicine, University Medical Centre Hamburg-Eppendorf
-
- Jacobi Janin
- Department of Diagnostic and Interventional Radiology and Nuclear Medicine, University Medical Centre Hamburg-Eppendorf
-
- Mathies Franziska
- Department of Diagnostic and Interventional Radiology and Nuclear Medicine, University Medical Centre Hamburg-Eppendorf
-
- Klene Christiane
- Department of Diagnostic and Interventional Radiology and Nuclear Medicine, University Medical Centre Hamburg-Eppendorf
-
- Zsebe Zsofia
- Department of Diagnostic and Interventional Radiology and Nuclear Medicine, University Medical Centre Hamburg-Eppendorf
-
- Klutmann Susanne
- Department of Diagnostic and Interventional Radiology and Nuclear Medicine, University Medical Centre Hamburg-Eppendorf
-
- Apostolova Ivayla
- Department of Diagnostic and Interventional Radiology and Nuclear Medicine, University Medical Centre Hamburg-Eppendorf
-
- Nakajima Kenichi
- Department of Functional Imaging and Artificial Intelligence, Kanazawa University Graduate School of Medicine
-
- Mester Janos
- Department of Diagnostic and Interventional Radiology and Nuclear Medicine, University Medical Centre Hamburg-Eppendorf
この論文をさがす
抄録
Background: In myocardial gated single-photon emission computed tomography (GSPECT), to differentiate true changes of left ventricular ejection fraction (LVEF) from inherent methodical variability is clinically relevant; however, data about repeatability of GSPECT LVEF in the same patients are rather inconsistent in literature. The aim of this study was therefore to determine repeatability coefficient (RC) of GSPECT LVEF at rest and to investigate the effect of the introduction of processing constraints in left ventricular edge detection.<br>Methods: Thirty-five patients referred for one-day myocardial GSPECT stress-rest scan were included. After the routine stress-rest study, patients were completely repositioned on the imaging table for a second rest acquisition using the same acquisition parameters. LVEF was computed using Corridor 4DM software without and with manual alignment of valve plane. Repeatability was assessed using the Bland-Altman method.<br>Results: RC of LVEF from unaligned datasets was 7.6% with upper and lower limits of agreement of 7.4% to -7.8%. After valve plane and ventricular long-axis length alignment, RC improved to 3.6% with upper and lower limits of agreement of 3.4% to -3.8%.<br>Conclusions: RC using unaligned determination of GSPECT LVEF was comparable to that from previous publications. However, RC using valve plane alignment could be improved to below 4% on 95% confidence level.
収録刊行物
-
- Annals of Nuclear Cardiology
-
Annals of Nuclear Cardiology 7 (1), 27-32, 2021-08-31
日本心臓核医学会
- Tweet
キーワード
詳細情報 詳細情報について
-
- CRID
- 1390289225019655552
-
- NII論文ID
- 130008081341
-
- ISSN
- 24241741
- 21893926
-
- 本文言語コード
- en
-
- データソース種別
-
- JaLC
- Crossref
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可