潅流下における犬心を用いた心室細動時の心磁図計測

書誌事項

タイトル別名
  • MCG Measurement during Ventricular Fibrillation Produced by an Isolated Canine Heart under Perfusion.

この論文をさがす

抄録

This study focuses on the magnetocardiogram (MCG) measurement of ventricular fibrillation (VF) for an isolated canine heart under perfusion. We carried out the MCG measurement in a prefabricated magnetically shielded room, using a 5-channel rf-SQUID connected to second- order gradiometers with base lines of 3.5cm and coil diam- eters of 2.5 cm. In a study of VF induced by electric shock, a fast Fourier transform (FFT) was used to compare the frequency characteristics of MCG and ECG recordings under VF with normal MCG and ECG. It was found that frequencies of 3 to 7 Hz dominated in both MCG and ECG recordings. Distribution of the magnetic field during VF showed a complicated pattern. In some periods, however, there was a characteristic pattern, which can be predicted to be a clockwise and/or counterclockwise circular current with a fundamental period of 0.2 s.

収録刊行物

被引用文献 (2)*注記

もっと見る

参考文献 (16)*注記

もっと見る

詳細情報 詳細情報について

  • CRID
    1390282680068771200
  • NII論文ID
    20000347074
  • NII書誌ID
    AN0031390X
  • DOI
    10.3379/jmsjmag.20.721
  • ISSN
    18804004
    02850192
    http://id.crossref.org/issn/02850192
  • 本文言語コード
    ja
  • データソース種別
    • JaLC
    • Crossref
    • CiNii Articles
  • 抄録ライセンスフラグ
    使用不可

問題の指摘

ページトップへ